Title of article :
Determination of urinary 8-hydroxy-2′-deoxyguanosine in obese patients by HPLC with electrochemical detection Original Research Article
Author/Authors :
Eva Samcov?، نويسنده , , Petr Marhol، نويسنده , , Franti?ek Opekar، نويسنده , , Jan Langmaier، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
A simple amperometric detector with a platinum tubular electrode was tested for the determination of urinary 8-hydroxy-2′-deoxyguanosine (8-OHdG) by high-performance liquid chromatography. Analyte was extracted from human urine using a single-step solid-phase extraction with the average recovery of 71±7%. Analysis was carried out on a C-18 analytical column under isocratic conditions. The limit of detection of 8-OHdG of 3 nM (S/N=3) was comparable with that achieved using the conventional amperometric detector. Procedure was used for the determination of 8-OHdG levels in 10 obese patients (body-mass index>30 kg m−2) yielding the mean concentration 32±27.1 nM of 8-OHdG or 1.85±0.72 μmol 8-OHdG mol−1 creatinine (4.63±1.8 μg 8-OHdG g−1 creatinine), which is not much different from the levels found in healthy individuals.
Keywords :
8-Hydroxy-2?-deoxyguanosine , DNA , Amperometric detector , Obese patient
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta