• Title of article

    Rapid determination of trace levels of tetracyclines in surface water using a continuous flow manifold coupled to a capillary electrophoresis system Original Research Article

  • Author/Authors

    L Nozal، نويسنده , , L Arce، نويسنده , , B.M Simonet، نويسنده , , A R??os، نويسنده , , M Valc?rcel، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    6
  • From page
    89
  • To page
    94
  • Abstract
    A method for the determination of three tetracyclines (viz. tetracycline, oxytetracycline and doxycycline) is proposed. The method uses a flow manifold coupled on-line to a capillary electrophoresis (CE) system. The tetracyclines are preconcentrated on a STRATA-X solid-phase minicolumn inserted in the manifold. This provides an average recovery of tetracyclines of 92% from spiked water samples. The proposed method allows the resolution of mixtures of tetracyclines at concentrations ranging from 7 to 45 g l−1 and has a limit of detection of 2 g l−1. Under optimal conditions, recoveries range from 99 to 106% and the precision is better than 7.5%. Based on the results, the method is a highly efficient choice for the separation and determination of tetracyclines in surface water samples.
  • Keywords
    Capillary electrophoresis , Tetracyclines , Preconcentration , STRATA-X , Environmental monitoring , Water analysis
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2004
  • Journal title
    Analytica Chimica Acta
  • Record number