Title of article :
Amperometric glucose biosensor based on sol–gel-derived metal oxide/Nafion composite films Original Research Article
Author/Authors :
Han Nim Choi، نويسنده , , Min Ah Kim، نويسنده , , Won-Yong Lee، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
9
From page :
179
To page :
187
Abstract :
Amperometric glucose biosensors have been developed based on the encapsulation of glucose oxidase in the nanoporous composite films of sol–gel-derived metal oxide (titania and silica) and perfluorosulfonated ionomer, Nafion, on a platinized glassy carbon electrode. The presence of Nafion polymer in the sol–gel-derived metal oxide not only prevents the cracking of the pure sol–gel-derived metal oxide film but also improves the sensitivity and long-term stability of the biosensor. Analytical performance of the glucose biosensor based on the titania/Nafion composite is superior to that of the biosensor based on the silica/Nafion composite in terms of response time, sensitivity, and long-term stability. The glucose biosensor based on the titania/Nafion composite film can reach 95% of steady-state current less than 2 s. The biosensor responds to glucose linearly up to 7 mM with a sensitivity of 15 μA/mM. The biosensor retained 80% of its initial activity after 4 months of storage in 0.05 M phosphate buffer at pH 7. Sol–gel-derived titania/Nafion composite can also be used for the preparation of second-generation (mediator-based) biosensor by immobilizing the positively charged mediator such as N,N′-dimethylaminomethyl ferrocene in the composite film via electrostatic and hydrophobic interactions.
Keywords :
Glucose , Sol–gel technique , Biosensor , Titania/Nafion composite , Silica/Nafion composite
Journal title :
Analytica Chimica Acta
Serial Year :
2005
Journal title :
Analytica Chimica Acta
Record number :
1030645
Link To Document :
بازگشت