• Title of article

    Time measurement-visual analysis of nickel(II) using autocatalytic reaction with sodium sulfite/hydrogen peroxide system and its application to the length detection-flow analysis Original Research Article

  • Author/Authors

    Jun Kato، نويسنده , , Tetsuo Higuchi، نويسنده , , Toshio Kubota، نويسنده , , Shukuro Igarashi، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    8
  • From page
    245
  • To page
    252
  • Abstract
    Trace amounts of nickel(II) can function as a trigger (=reaction initiator) in an autocatalytic reaction with the sodium sulfite/hydrogen peroxide system. Based on this finding, sub-μg L−1 levels of nickel(II) were determined by a time measurement using the autocatalytic reaction. The detection range using the above method was 10−9–10−5 M, the detection limit (3σ) was 8.1 × 10−10 M (0.047 μg L−1), and the relative standard deviation was 2.66% at nickel(II) concentration of 10−7 M (n = 7). This method was applied to length detection-flow injection analysis. The detection range for the flow injection analysis was 2 × 10−9–2 × 10−3 M. The detection limit (3σ) was 1.4 × 10−9 M (0.082 μg L−1), and the relative standard deviation was 1.86 at initial nickel(II) concentration of 10−6 M (n = 7).
  • Keywords
    Length detection , visual analysis , Autocatalytic reaction , nickel , Time measurement , Flow injection analysis
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2007
  • Journal title
    Analytica Chimica Acta
  • Record number

    1030830