Title of article
Assessment of the uncertainty budget for the amperometric measurement of dissolved oxygen Original Research Article
Author/Authors
Paola Fisicaro، نويسنده , , Annemie Adriaens، نويسنده , , Enzo Ferrara، نويسنده , , Enrico Prenesti، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
7
From page
75
To page
81
Abstract
This work aimed at identifying the main sources of uncertainty for the measurement of dissolved oxygen concentration in aqueous solutions. The experimental apparatus consists of an amperometric cell based on the Clark-type sensor. The corresponding uncertainty budget was assessed, this being a fundamental step for the validation of a measurement method. The principle of the measurement, as well as the procedure for the set-up and the characterisation of the cell, are described. The measurement equation was defined as a combination of Faradayʹs and Fickʹs laws, and a method was worked out for the empirical determination of the diffusivity parameter. In this connection, the solutions of oxygen were standardised by way of the Winklerʹs titration, as suggested by the ISO Guide 5813 and 5814. With this approach we aimed at contributing to the development of a potential primary method of measurement. A discussion of all the contributions to the overall uncertainty is reported, allowing operators to locate the largest ones and plan specific improvements.
Keywords
Clark-type sensors , Dissolved oxygen measurement uncertainty , validation , Amperometric measurement
Journal title
Analytica Chimica Acta
Serial Year
2007
Journal title
Analytica Chimica Acta
Record number
1031059
Link To Document