Title of article :
Potentiometric flow-injection system for determination of alkaline phosphatase in human serum Original Research Article
Author/Authors :
Dominika Ogo?czyk، نويسنده , , Robert Koncki، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
A flow-injection system for detection of alkaline phosphatase (ALP) activity in human serum samples has been developed. As a specific and inexpensive ALP substrate for this kinetic assay monofluorophosphate (MFP) was applied. For detection of fluoride ions, generated in the course of the biocatalytic hydrolysis of MFP, conventional fluoride ion-selective electrode based on LaF3-crystalline membrane was applied. After optimization the system allows analysis of human serum with high selectivity and relatively short time of analysis (5–6 samples h−1). Volume of serum required for analysis is 0.05 mL. The system is useful for determination of the enzyme activity in human serum samples at physiological and pathological levels as well as for detection of isoenzymatic forms of ALP.
Keywords :
Human serum , Flow-injection analysis , Potentiometry , Alkaline phosphatase
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta