Title of article :
Simultaneous voltammetric detection of ascorbic acid, dopamine and uric acid using a pyrolytic graphite electrode modified into dopamine solution Original Research Article
Author/Authors :
Robson P. da Silva، نويسنده , , Antonio William O. Lima، نويسنده , , Silvia H.P. Serrano، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
Pyrolytic graphite electrodes (PGE) were modified into dopamine solutions using phosphate buffer solutions, pH 10 and 6.5, as supporting electrolyte. The modification process involved a previous anodization of the working electrode at +1.5 V into 0.1 mol L−1 NaOH followed by other anodization step, in the same experimental conditions, into dopamine (DA) solutions. pH of the supporting electrolyte performed an important role in the production of a superficial melanin polymeric film, which permitted the simultaneous detection of ascorbic acid (AA), (DA) and uric acid (UA), ΔEAA–DA = 222 mV; ΔEAA–UA = 360 mV and ΔEDA–UA = 138 mV, avoiding the superficial poisoning effects. The calculated detection limits were: 1.4 × 10−6 mol L−1 for uric acid, 1.3 × 10−5 mol L−1 for ascorbic acid and 1.1 × 10−7 mol L−1 for dopamine, with sensitivities of (7.7 ± 0.5), (0.061 ± 0.001) and (9.5 ± 0.05) A mol−1 cm−2, respectively, with no mutual interference. Uric acid was determined in urine, blood and serum human samples after dilution in phosphate buffer and no additional sample pre-treatment was necessary. The concentration of uric acid in urine was higher than the values found in blood and serum and the recovery tests (92–102%) indicated that no matrix effects were observed.
Keywords :
Pyrolytic graphite modified electrode , Dopamine , Uric and ascorbic acids , Biological fluids , Differential pulse voltammetry
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta