• Title of article

    Determination of daunorubicin in human urine by capillary zone electrophoresis with amperometric detection Original Research Article

  • Author/Authors

    Qin Hu، نويسنده , , Lan Zhang، نويسنده , , Tianshu Zhou، نويسنده , , Yuzhi Fang، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    5
  • From page
    15
  • To page
    19
  • Abstract
    A simple, reliable and reproducible method, based on capillary zone electrophoresis with amperometric detection, for the determination of daunorubicin in human urine is described in this paper. A carbon disk electrode used as working electrode. The optimal conditions of separation and detection were pH 7.8 phosphate buffer (0.15 mol/l), 8 kV for the separation voltage and 0.95 V (versus Ag/AgCl, 3 mol/l KCl) for the detection potential. The linear range was from 2×10−6 to 2×10−4 mol/l with the regression coefficient of 0.9995, and the detection limit was 8×10−7 mol/l. The method built in this paper was directly applied to the determination of daunorubicin in the imitation human urine without any other sample pre-treatment except filtration, and the assay result was satisfactory.
  • Keywords
    Anthracycline , Capillary zone electrophoresis , Daunorubicin , Electrochemical detection
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2000
  • Journal title
    Analytica Chimica Acta
  • Record number

    1031863