• Title of article

    Adsorption voltammetry of the mix-polynuclear complex of zirconium–calcium–alizarin red S at a carbon paste electrode Original Research Article

  • Author/Authors

    Ju-Nan Li، نويسنده , , Jun Zhang، نويسنده , , Pei-Hong Deng، نويسنده , , Ya-Qing Peng، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    7
  • From page
    81
  • To page
    87
  • Abstract
    A novel new method is described for the determination of zirconium based on the cathodic adsorptive stripping of the mix-polynuclear complex of zirconium–calcium–alizarin red S (ARS) at a carbon paste electrode (CPE). The second-order derivative linear scan voltammograms of the adsorbed mix-polynuclear complex of zirconium–calcium–ARS are recorded by model JP-303 polarographic analyzer from +0.20 to 0.80 V (versus SCE). Optimal conditions are found to be: a glycine (0.096 M)–potassium biphthalate (0.036 M) buffer solution (pH 4.0) containing 8.0×10−4 M Ca2+ and 2.0×10−5 M ARS; accumulation potential, +0.20 V; accumulation time, 90 s; rest time, 10 s; scan rate, 500 mV s−1. The results show that the complex can be adsorbed on the surface of the CPE, yielding one peak at −0.50 V, corresponding to reduction of ARS in the complex at the electrode. The detection limit is found to be 4.0×10−11 M with 4 min of accumulation time. The linear range is from 1.0×10−10 to 2.0×10−7 M. Application of the procedure to the determination of zirconium in the ore samples gives good results.
  • Keywords
    Adsorption voltammetry , Carbon paste electrode , Zirconium , Mix-polynuclear complex
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2001