Title of article :
Determination of ultratraces of nitrite by solid-phase preconcentration using a novel flow-through spectrophotometric optrode Original Research Article
Author/Authors :
Manuel Mir?، نويسنده , , Wolfgang Frenzel، نويسنده , , Victor Cerda، نويسنده , , José Manuel Estela، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Abstract :
A novel sorbent-packed planar wall microcolumn-based optosensor coupled to a flow injection analysis system (FIA) for the determination of traces of nitrite by solid-phase spectrophotometry is presented. The methodology is based on the on-line nitrite derivatization with a modified Griess reagent to form an azo-dye whose retention onto octadecyl-covalently bonded silica gel by physical adsorption and partitioning is continuously monitored. After the analytical signal has been recorded, the sorbed compound is eluted with 80% methanol/water (v/v), rendering the system ready for the next extraction.
Selection of the solid support and the suitable flow-through cell configuration are discussed in detail. By minimization of the blank signal, a concentration of nitrite as low as 1 ng/ml can be easily determined using 2.5 ml of sample. Detection limit (3σblank) of 0.46 ng/ml of nitrite, relative standard deviation better than 4%, sample throughput of 12 h−1 and an enrichment factor of 93 in relation to the conventional spectrophotometric method were obtained.
The utility of the present optosensing system, wherein a mass calibration is also feasible, was testified by the satisfactory results obtained in its application to tap, rain and aquarium water samples.
Keywords :
Optosensing , Nitrite , Preconcentration , Solid-phase extraction
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta