• Title of article

    Flow injection analysis of the insecticide imidacloprid in water samples with photochemically induced fluorescence detection Original Research Article

  • Author/Authors

    J.L. Vilchez، نويسنده , , M.C Valencia، نويسنده , , A Naval?n، نويسنده , , Molinero-Morales، B. نويسنده , , L.F Capit?n-Vallvey، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    7
  • From page
    299
  • To page
    305
  • Abstract
    A flow injection analysis (FIA) system, combined with photochemical induced fluorescence (PIF) detection is developed for the sensitive and rapid determination of imidacloprid. It is based on the conversion of imidacloprid into the fluorophore 1-(6-chloro-3-pyridyl-methyl)-2-(hydroxyimino)-3,4-didehydroimidalozolidene. In an aqueous medium, this compound shows native fluorescence with an excitation maximum at 334 nm and an emission maximum at 377 nm. The linear concentration range of application was 1.0–60.0 ng ml−1 of imidacloprid, with a relative standard deviation of 2.1% (for a level of 10 ng ml−1) and a detection limit of 0.3 ng ml−1. The method was applied to check whether imidacloprid was present above this limit in waters from Cuenca and Granada (Spain). It was validated applying a recovery test (Student’s t-test). Recovery levels of the method reached around 100% in all cases.
  • Keywords
    Water analysis , Imidacloprid , Photochemically induced fluorescence (PIF) , Flow injection analysis (FIA)
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2001
  • Journal title
    Analytica Chimica Acta
  • Record number

    1032511