Title of article
Development of a heterogeneous chemiluminescent flow immunoassay for DDT and related compounds Original Research Article
Author/Authors
Alexandra E. Botchkareva، نويسنده , , Fabiana Fini، نويسنده , , Sergei Eremin، نويسنده , , Josep V. Mercader، نويسنده , , Angel Montoya، نويسنده , , Stefano Girotti، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2002
Pages
10
From page
43
To page
52
Abstract
A heterogeneous chemiluminescent (CL) flow immunoassay for DDT was optimized comparing different types of immunoaffinity supports: beads, nylon coils and membranes (membranes HyBondN+). In order to characterize solid immunoaffinity supports two basic immunoassay formats were performed, using (1) enzyme-labeled secondary and (2) enzyme-labeled specific monoclonal antibodies (MAbs). In both formats, hapten DDT5 conjugated to ovalbumin immobilized on solid supports according to the appropriate immobilization procedure, enzyme label (horseradish peroxidase, HRP) and luminescent detection (luminol/H2O2/p-iodophenol) were used. The lowest limit of detection (LOD), 1 nM p,p-DDT, was obtained with a membrane-based flow immunoassay with HRP-labeled specific antibody. Beads and packed tubing were discarded as appropriate supports because of the difficulties encountered for reproducible packing and the occurrence of light scatterring (beads), which seriously compromised the performance and reproducibility of the flow immunoassay.
Keywords
Enhanced chemiluminescence , Flow immunoassay , Immunosupports , Pesticides