Title of article :
Chemiluminescence multichannel immunosensor for biodetection Original Research Article
Author/Authors :
Erwin Yacoub-George، نويسنده , , Leonhard Meixner، نويسنده , , Waltraud Scheithauer، نويسنده , , Andreas Koppi، نويسنده , , Stephan Drost، نويسنده , , Hans Wolf، نويسنده , , Christine Danapel، نويسنده , , Klaus A Feller، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
10
From page :
3
To page :
12
Abstract :
An improved portable detector for biological compounds, the chemiluminescence multichannel immunosensor (CL-MADAG), has been developed and characterised. The device is based on a capillary ELISA technique in combination with a miniaturised fluidics system and uses chemiluminescence as the detection principle. The fluidics system construction allows three chemiluminescence immunoassays to be performed simultaneously within three fused silica capillaries (FSC). The CL-MADAG was characterised in a series of experiments with staphylococcal enterotoxin B (SEB) as a model toxin, the bacterial phage virus M13 as a virus simulant, and a pathogenic strain of Escherichia coli as simulant for bacteria. It was shown that the CL-MADAG can assay liquid samples for these substances within 24 min. The detection limits were 5 ng/ml for SEB, 105 cfu/ml for E. coli O157:H7 and 107 pfu/ml for M13.
Keywords :
Biosensor , Immunoassay , Multianalyte detection , Chemiluminescence