  • Abstract
    A calibration method has been developed which is realised in the flow injection analysis (FIA) by the gradient technique. According to this method two transient peaks, one for a sample and the other for a sample with standard addition, are recorded and compared point by point in the entire defined time range. The analytical result is estimated on the basis of information gained about the local analyte concentrations in the sample zone. The method allows the results to be reliable when both the non-linear calibration dependence and the interference effect occur. As an example, calcium in synthetic samples containing silicon, phosphate, aluminium, vanadium and titanium, and also in iron ore sample, were determined by flame atomic absorption spectrometry (FAAS). It has been proved, that the method can be effective in overcoming even extremely strong interferences, providing analytical results with average accuracy equal to ca. 5% and with precision 2–3 times inferior to that obtained by conventional FI calibration.
  • Keywords
    Ion exchange , pH , calcium , Diffusion , Magnesium , Diffusive Gradients in Thin Films