Abstract :
Determination of small amounts of the fat-soluble species Vitamin A (VA) (2.5 μg/g) and β-carotene (9 μg/g) from emulsified nutritional supplements containing 50 kinds of co-existing compounds and a fat content between 2000 and 8000 times higher was performed by solid-phase extraction (SPE) and high-performance liquid chromatography (HPLC) with fluorescence detection set at ex. 350 nm and em. 480 nm, and visible detection at 450 nm using an Inertsil ODS 80A (5 μm) analytical column. Mobile phases of methanol–ethanol (50:50) and acetonitrile–ethanol (70:30) were used for the both vitamins. A Bond Elut C18 cartridge was chosen for SPE after comparison with eight other types of SPE cartridge. Retention time of VA and β-carotene was 7 and 8 min, respectively, giving a limit of detection of ca. 0.1 ng per injection at a signal-to-noise ratio 3:1. Recoveries of VA and β-carotene were over 90% by the standard addition method. Relative standard deviation of VA and β-carotene were ca. 2.9 (n=5) and 2.3% (n=5), respectively.