Title of article :
Nickel oxide-modified composite electrode for electrochemical detection of polyhydroxyl compounds in liquid chromatographic analysis Original Research Article
Author/Authors :
Hyekyoung Chung، نويسنده , , Heejin Yang، نويسنده , , Won Kim، نويسنده , , Jongman Park، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Abstract :
Nickel oxide-modified polymeric composite electrode was prepared for electrochemical detection in liquid chromatography (LC-ECD) or flow injection analysis of polyhydroxyl compounds. The composite electrode was prepared by incorporating semigraphitic carbon black and nickel oxide particles in polystyrene polymer matrix cross-linked with divinylbenzene. The analytical characteristics of the electrodes in LC-ECD and flow injection analysis of polyhydroxyl compounds were evaluated in terms of sensitivity, reproducibility, stability and surface renewability. The electrode revealed three–four orders of dynamic concentration ranges from sub-micromolar to 10−3 M with high correlation coefficients from 0.9998 to 0.9999 for various polyhydroxyl compounds. The standard deviation for the surface restoration was <2%. High reproducibility, stability, and the renewability of surface are thought to be due to the catalyst particles highly dispersed through robust polymeric electrode matrix.
Keywords :
Electrochemical detection , Polyhydroxyl compounds , Liquid chromatographic analysis
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta