• Title of article

    Development of a method for estimating an accurate equivalence point in nickel titration of cyanide ions Original Research Article

  • Author/Authors

    Toshihiro Suzuki، نويسنده , , Akiharu Hioki، نويسنده , , Masayasu Kurahashi، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    7
  • From page
    159
  • To page
    165
  • Abstract
    In a nickel titration of cyanide ions using murexide as indicator, an accurate equivalence point was determined by a non-linear least-squares curve-fitting for a titration curve. This method was developed to establish a standard solution for cyanide ions. In a curve-fitting procedure, a theoretical titration curve was calculated, assuming that nickel ion formed only a 1:4 Ni2+:CN− complex with cyanide ions and formed only a 1:1 complex with murexide. Results of the curve-fitting were reasonable at any pH and any indicator concentration studied. The combined standard uncertainty for a concentration of a 1000 mg kg−1 cyanide solution by this method was 0.079%.
  • Keywords
    Complexometric titration , Equivalence point , Murexide , Uncertainty , Nickel ion , Cyanide ion
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2003
  • Journal title
    Analytica Chimica Acta
  • Record number