Title of article :
Application of an integrated microchip system with capillary array electrophoresis to optimization of enzymatic reactions Original Research Article
Author/Authors :
Joon Myong Song، نويسنده , , Guy D. Griffin، نويسنده , , Tuan Vo-Dinh، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
In this work, a combination of complementary metal-oxide semiconductor (CMOS) microchip system with capillary array electrophoresis (CAE) is demonstrated as a system for optimizing conditions for enzymatic reaction. Dimethylacridinone (DDAO)-phosphate substrate and alkaline phosphatase conjugate were selected for the enzymatic reaction, which was applicable to the enzyme-linked immunosorbent assay (ELISA) technique. Laser-induced fluorometry with a miniature semiconductor laser was used to detect the enzymatic products. The speed of the enzymatic reaction between the DDAO-phosphate and the alkaline phosphatase conjugate was investigated as a function of reaction time. The microchip-CAE detection system could determine the pH condition and the concentration of enzyme that are suitable for rapid and low-cost analysis. This result shows the feasibility of using the microchip-CAE system for application to miniaturized screening systems.
Keywords :
Capillary array electrophoresis , Complementary metal-oxide semiconductor microchip , Enzymatic reaction
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta