Title of article :
Determination of cadmium in human urine by electrothermal atomic absorption spectrometry Original Research Article
Author/Authors :
Cl?udio José de Castro Maciel، نويسنده , , Gilmar Machado Miranda، نويسنده , , Danielle Palma de Oliveira، نويسنده , , Maria Elisa P.B. de Siqueira، نويسنده , , Josianne Nic?cio Silveira، نويسنده , , Edna Maria Alvarez Leite، نويسنده , , José Bento Borba da Silva، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
7
From page :
231
To page :
237
Abstract :
In the present work, a fast acid digestion method was applied to the determination of cadmium in urine samples by electrothermal atomic absorption spectrometry (ET AAS). The pyrolysis and atomization temperatures were 400 and 1200 °C with a characteristic mass of 1.3 pg. Quantification was made with matrix matching calibration curves. The method was validated by the analysis of certified urine samples. Differences were not observed among the levels of the metal in populations from metropolitan and rural areas. The limit of detection (LOD, k=3, n=10) was 0.08 μg l−1. Coefficients of variation for intra- and interassay studies were 5.5 and 12.7%, respectively. Recovery of spiked samples showed acceptable values. Significant correlation was observed among sex, with higher values for women. Age also showed an influence on urinary Cd levels, being higher for individuals over 46 years. No correlation was found between either tabagism or alcoholic beverage intake and of urinary cadmium levels.
Keywords :
Human urine , Electrothermal atomic absorption spectrometry , cadmium
Journal title :
Analytica Chimica Acta
Serial Year :
2003
Journal title :
Analytica Chimica Acta
Record number :
1033672
Link To Document :
بازگشت