    Improving the lower detection limit of potentiometric sensors by covalently binding the ionophore to a polymer backbone Original Research Article

  • Author/Authors

    Martin Püntener، نويسنده , , Tam?s Vigassy، نويسنده , , Ellen Baier، نويسنده , , Alan Ceresa، نويسنده , , Ern? Pretsch، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    8
  • From page
    187
  • To page
    194
  • Abstract
    The lower detection limit of polymeric membrane ion-selective electrodes (ISEs) is impaired by zero-current ion fluxes through the organic phase. This adverse effect is largely eliminated by covalently attaching the ionophore to a polymer backbone. To this purpose, the Pb2+-selective ligand, 4-tert-butylcalix[4] arene-tetrakis(N,N′-dimethylthioacetamide) is substituted on its upper rim by a diol derivative which is subsequently copolymerized with poly(tetrahydrofuran)diol and 2,2,4-trimethylhexamethylene diisocyanate to the corresponding polyurethane. By measurements on sandwich membranes, it is shown that through binding the ionophore to the polymer, the mobility of Pb2+ in the ISE membrane is strongly reduced. As a consequence, the response range of such an ISE is extended by several orders of magnitude. This is the case even when using an internal electrolyte that with an ISE based on a mobile ionophore leads to strong deviations from the linear response because of ion uptake from the sample into the membrane or ion release from the membrane into the sample. With a conventional inner filling solution of 10−1 M Pb(NO3)2, a lower detection limit of 1.7×10−9 M Pb2+ has been achieved in the presence of 10−4 M Na+.
  • Keywords
    Potentiometric sensor , Detection limit , Polymeric ionophore , Ion fluxes
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2004
  • Journal title
    Analytica Chimica Acta
  • Record number

    1033763