Title of article :
Simultaneous determination of pH, chloride and nickel in electroplating baths using sequential injection analysis Original Research Article
Author/Authors :
José Edson da Silva، نويسنده , , Maria Fernanda Pimentel، نويسنده , , Valdinete Lins da Silva، نويسنده , , Maria da Conceiç?o B.S.M Montenegro، نويسنده , , Alberto N. Araujo، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
A sequential injection analysis system was developed to quantify pH, chloride and nickel in electrolytic baths, in the ranges 1–5 pH units, and 0.1–1.0 and 0.1–1.6 mol l−1, respectively. To enable pH and chloride determination, potentiometric detection with two ion-selective electrodes in a tubular configuration was used. Nickel concentrations were assessed using colorimetric detection at 660 nm. pH was determined prior to nickel determination and just after sample injection (500 μl) into a 0.025 mol l−1 phosphate buffer carrier stream at pH 6.3 and a 9.10 ml min−1 flow rate. For chloride determination, on-line dialysis through a cellulose membrane was used to enable sample dilution and matrix separation. A 25−1 fractional factorial design based on the carrier solution composition and the levels of the hydrodynamic parameters was used for system optimization. At the optimized settings a sampling rate of 40 samples h−1 was attained, with precision and accuracy statistically indistinguishable from those achieved with conventional procedures.
Keywords :
Electroplating bath , nickel , pH and chloride , Sequential injection analysis with potentiometric , Spectrophotometric detection
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta