    Determination of trace lead in the high salt matrix of comestible CaCO3 by an isotope dilution method with detection by inductively coupled plasma-mass spectrometry Original Research Article

  • Author/Authors

    Zhiyong Huang، نويسنده , , Chao-Yong Yang، نويسنده , , Zhi-Xia Zhuang، نويسنده , , Xiao-Ru Wang، نويسنده , , Frank S.C. Lee، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    6
  • From page
    77
  • To page
    82
  • Abstract
    The determination of trace lead in a CaCO3 matrix by isotope dilution coupled anion-exchange and quadrupole-based inductively coupled plasma-mass spectrometry was established. The matrix of CaCO3 seriously affected the determination by clogging the cones of the instrument was removed by the anion-exchange resin column. Effects such as dead time, mass bias, dwell time and the amount of isotopic agent spiked were discussed and corrected. The detection limit estimated by 3× standard deviation of the procedure blank was 2.94 ng g−1. The results indicated that the lead concentration in all samples tested was far less than 3%, the legislated limit. The recovery mostly ranged from 90 to 110%. The relative standard deviation was found to be <1%.
  • Keywords
    Isotope dilution , Anion-exchange , CaCO3 , Quadrupole-based inductively coupled plasma-mass spectrometry , lead
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2004
  • Journal title
    Analytica Chimica Acta
  • Record number