Title of article :
Detection limits and goodness-of-fit measures for the two-component model of chemical analytical error Original Research Article
Author/Authors :
Machelle D. Wilson*، نويسنده , , David M. Rocke، نويسنده , , Blythe Durbin، نويسنده , , Henry D. Kahn، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
The utility of analytical chemistry measurements in most applications is dependent on an assessment of measurement error. This paper demonstrates the use of a two-component error model in setting limits of detection and related concepts and introduces two goodness-of-fit statistics for assessing the appropriateness of the model for the data at hand. The model is applicable to analytical methods in which high concentrations are measured with approximately constant relative standard deviation. At low levels, the relative standard deviation cannot stay constant, since this implies vanishingly small absolute standard deviation. The two-component model has approximately constant standard deviation near zero concentration, and approximately constant relative standard deviation at high concentrations, a pattern that is frequently observed in practice. Here we discuss several important applications of the model to environmental monitoring and also introduce two goodness-of-fit statistics, to ascertain whether the data exhibit the error structure assumed by the model, as well as to look for problems with experimental design.
Keywords :
Two-component error model , Chemical analytical error , Goodness-of-fit , Limits of detection
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta