Abstract
A new and simple flow injection system procedure has been developed for the indirect determination of cyanide. The method is based on insertion of aqueous cyanide solutions into an on-line cadmium carbonate packed column (25% m/m suspended on silica gel beads) and a sodium hydroxide with pH 10 is used as the carrier stream. The eluent containing the analyte as cadmiumcyanide complexes, produced from reaction between cadmium carbonate and cyanide, measured by flame atomic absorption spectrometry. The absorbance is proportional to the concentration of cyanide in the sample. The linear range of the system is up to 15 mg L−1 with a detection limit 0.2 mg L−1 and sampling rate 72 h−1. The method is suitable for determination of cyanide in industrial waste waters with a relative standard deviation better than 1.22%.
Keywords
Flame atomic absorption spectrometry , Indirect determination , Cyanide , Cadmium carbonate , Flow injection analysis