• Title of article

    Miniaturised stripping-based carbon modified sensor for in field analysis of heavy metals Original Research Article

  • Author/Authors

    Ilaria Palchetti، نويسنده , , Serena Laschi، نويسنده , , Marco Mascini، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    7
  • From page
    61
  • To page
    67
  • Abstract
    A new miniaturised electrochemical carbon modified sensor for on-site stripping analysis of trace heavy metals is described. The graphite working electrode is modified with a cellulose-derivative mercury coating. A detailed description of the construction of the system as well as the systematic laboratory test carried out to optimise its performance is reported. Coupling this sensor with square wave (SW) anodic stripping analysis, detection limits of 0.3, 1 and 0.5 μg l−1 were found for lead(II), cadmium(II) and copper(II), respectively. Deoxygenation of sample solution is not required. The entire analysis time is shorter than 4 min. Examples of on-site environmental applications are given. In particular, on-site analysis of heavy metals was performed at a contaminated site. For this purpose, the sensor was coupled with a portable commercial electroanalytical instrumentation.
  • Keywords
    Soil , sediments , Stripping voltammetry , Mercury-coating , Screen-printed electrodes , heavy metals , wastewater , On-site analysis
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2005
  • Journal title
    Analytica Chimica Acta
  • Record number

    1034492