• Title of article

    High-performance liquid chromatography determination of triptolide in vitro permeation studies Original Research Article

  • Author/Authors

    Xue-Ling Chang، نويسنده , , Huabing Chen، نويسنده , , Xiao-Zhi Zhao، نويسنده , , Zhong-Hong Gao، نويسنده , , Huibi Xu، نويسنده , , Xiang-Liang Yang، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    7
  • From page
    215
  • To page
    221
  • Abstract
    A simple, rapid and sensitive high-performance liquid chromatography (HPLC) method has been developed for the determination of triptolide. Triptolide was separated from skin endogenous and blank matrices on a 5 μm LiChrospher RP-C18 column by a mobile phase of methanol–water (65:35, v/v). The permeation samples were injected directly without pretreatment. The limit of quantitation (LOQ) and detection (LOD) for triptolide in permeation samples were far below (0.01 and 0.005 μg/mL, respectively). The method was linear over the range of 0.1–104.2 μg/mL with r2 = 0.9999. This HPLC assay is promising for measuring in vitro percutaneous penetration of triptolide through mice skins and also can be performed in the triptolide-loaded microemulsions formulation screening.
  • Keywords
    Triptolide , Permeation study , HPLC
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2005
  • Journal title
    Analytica Chimica Acta
  • Record number

    1034637