Title of article :
Determination of creatinine in clinical samples based on flow-through microsystem Original Research Article
Author/Authors :
Ilona Grabowska، نويسنده , , Michal Chudy، نويسنده , , Artur Dybko، نويسنده , , Zbigniew Brzozka، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
5
From page :
181
To page :
185
Abstract :
The paper describes a flow-through microsystem with optical detection for measurement of creatinine concentration in urine and post-dialysate fluid. The microsystem is made of a transparent polymer-poly(dimethylosiloxane) (PDMS). In order to determine creatinine, classical Jaffe reaction was carried out. The influence of reaction parameters (temperature and solutions flow rate) on its sensitivity and repeatability of recorded signal was investigated. Results of creatinine determination in postdialysate fluids and urine were compared to results obtained in hospital laboratories.
Keywords :
Creatinine determination , Flow-through microsystem , Optical detection , Jaffe reaction , PDMS
Journal title :
Analytica Chimica Acta
Serial Year :
2005
Journal title :
Analytica Chimica Acta
Record number :
1034676