Title of article
Graphite impurities cause the observed ‘electrocatalysis’ seen at C60 modified glassy carbon electrodes in respect of the oxidation of l-cysteine Original Research Article
Author/Authors
Roohollah Torabi Kachoosangi، نويسنده , , Craig E. Banks، نويسنده , , Richard G. Compton، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
4
From page
1
To page
4
Abstract
The recently reported claims of electrocatalysis using C60 film modified glassy carbon electrodes [W.T. Tan, A.M. Bond, S.W. Ngooi, E.B. Lim, J.K. Goh, Anal. Chim. Acta 491 (2003) 181] for the oxidation of l-cysteine are questioned. We show that C60 itself is not electrocatalytic at the potentials concerned but rather it is likely that graphite impurities in the C60 material used by Tan et al. that provide their observed ‘electrocatalysis’.
Keywords
Graphite impurities , Electrocatalysis , l-Cysteine , C60 modified electrodes
Journal title
Analytica Chimica Acta
Serial Year
2006
Journal title
Analytica Chimica Acta
Record number
1035821
Link To Document