On-line monitoring of μg/L levels of haloacetic acids using ion chromatography with post-column nicotinamide reaction and fluorescence detection Original Research Article
Author/Authors
Paul S. Simone Jr.، نويسنده , , Greg T. Anderson، نويسنده , , Gary L. Emmert، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
8
From page
259
To page
266
Abstract
A method for measuring the concentrations of the five regulated haloacetic acids (HAA5) in drinking water is reported. This method uses ion chromatography to separate HAA5 species, followed by post-column reaction with nicotinamide and detection of the fluorescent products. The result of method detection limit, accuracy, precision, linearity and interference studies are reported. The on-line monitoring method is compared directly to USEPA 552.3 in Memphis drinking water. Though not meant to replace the USEPA 552.3 for compliance monitoring, the proposed method does offer attractive alternatives considering the ease of automation and application of on-line monitoring directly from drinking water distribution systems.
Keywords
Drinking water , Haloacetic acids (HAA5) , Water analysis , Ion chromatography , On-line monitoring , Disinfection by-products
Journal title
Analytica Chimica Acta
Serial Year
2006
Journal title
Analytica Chimica Acta
Record number
1035931
Link To Document