Title of article
Palladium and the electrochemical quartz crystal microbalance: A new method for the in situ analysis of the precious metal in aqueous solutions Original Research Article
Author/Authors
Nathan A. Carrington، نويسنده , , D. Lynn Rodman، نويسنده , , Zi-Ling Xue، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
6
From page
303
To page
308
Abstract
A new method for the quantitative determination of palladium(II) by the electrochemical quartz crystal microbalance (EQCM) technique has been developed. Using a bare carbon-coated quartz crystal, Pd(II) ions are directly deposited from aqueous solution as palladium metal onto the crystal surface, and the Pd(II) concentration is determined with a detection limit of 0.0156 mM, or 1.66 ppm. No complexing agent or preconcentration of palladium is required for the analysis. The palladium is stripped from the crystal through its electrochemical oxidation, regenerating the crystal for subsequent multi-cycle palladium analyses. A conventional gold-coated quartz crystal was incapable of carrying out the same measurements. The EQCM technique presented is simple, sensitive, and reproducible for the detection of this widely used precious metal.
Keywords
Electrochemical quartz crystal microbalance , Palladium , Carbon-coated quartz crystal , Electrochemistry