Title of article
Flow injection conductometric system with gas diffusion separation for the determination of Kjeldahl nitrogen in milk and chicken meat Original Research Article
Author/Authors
Jaroon Junsomboon، نويسنده , , Jaroon Jakmunee، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
7
From page
232
To page
238
Abstract
A simple flow injection (FI) conductometric system with gas diffusion separation was developed for the determination of Kjeldahl nitrogen (or proteins) in milk and chicken meat. The sample was digested according to the Kjeldahl standard method and the digest was diluted and directly injected into the donor stream consisting of 4 M NaOH. In alkaline medium, ammonium was converted to ammonia, which diffused through the PTFE membrane to dissolve in an acceptor stream (water). Dissociation of ammonia caused a change in conductance of the acceptor solution, which was linearly proportional to the concentration of ammonium originally present in the injected solution. A conductometric flow through cell and an amplifier circuit was fabricated, which helped improve sensitivity of the conductometric detection system. With using a plumbing Teflon tape as a gas diffusion membrane and without thermostating control of the system, a linear calibration graph in range of 10–100 mg L−1 N-NH4 was obtained, with detection limit of 1 mg L−1 and good precision (relative standard deviation of 0.3% for 11 replicate injections of 50 mg L−1 N-NH4). The developed method was validated by the standard Kjeldahl distillation/titration method for the analysis of milk and chicken meat samples. The proposed system had sample throughput of 35 h−1 and consumed much smaller amounts of chemical than the standard method (275 mg vs 17.5 g of NaOH per analysis, respectively).
Keywords
Flow injection , Gas diffusion , Conductometry , Proteins , Ammonium , Chicken , Milk , Kjeldahl
Journal title
Analytica Chimica Acta
Serial Year
2008
Journal title
Analytica Chimica Acta
Record number
1036496
Link To Document