Title of article
Determination of antimony(III) and total antimony by single-drop microextraction combined with electrothermal atomic absorption spectrometry Original Research Article
Author/Authors
Zhefeng Fan، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
5
From page
300
To page
304
Abstract
A simple and sensitive method for using electrothermal atomic absorption spectrometry (ET AAS) with Rh as permanent modifier determination of Sb(III) and total Sb after separation and preconcentration by N-benzoyl-N-phenylhydroxylamine (BPHA)–chloroform single drop has been developed. Parameters, such as pyrolysis and atomization temperature, solvent type, pH, BPHA concentration, extraction time, drop size, stirring rate and sample volume were investigated. Under the optimized experimental conditions, the detection limits (3σ) were 8.0 ng L−1 for Sb(III) and 9.2 ng L−1 for total Sb, respectively. The relative standard deviations (R.S.Ds.) were 6.6% for Sb(III) and 7.1% for total Sb (c = 0.2 ng mL−1, n = 7), respectively. The enrichment factor was 96. The developed method has been applied successfully to the determination of Sb(III) and total Sb in natural water samples.
Keywords
Electrothermal atomic absorption spectrometry , Single-drop microextraction , Permanent modifier , Antimony species
Journal title
Analytica Chimica Acta
Serial Year
2007
Journal title
Analytica Chimica Acta
Record number
1036937
Link To Document