Title of article :
Micro-Winkler titration method for dissolved oxygen concentration measurement Original Research Article
Author/Authors :
Irja Helm، نويسنده , , Lauri Jalukse، نويسنده , , Martin Vilbaste، نويسنده , , Ivo Leito، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
In this report a gravimetric micro-Winkler titration method for determination of dissolved oxygen concentration in water is presented. Mathematical model of the method taking into account all influence factors is derived and an uncertainty analysis is carried out to determine the uncertainty contributions of all influence factors. The method is highly accurate: the relative expanded uncertainties (k = 2) are around 1% in the case of small (9–10 g) water samples. The uncertainty analysis carried out in characterizing the uncertainty of the method is the most comprehensive published for a micro-Winkler method, resulting in experimentally obtained estimates for all uncertainty sources of practical significance (around 20 uncertainty sources altogether).
Keywords :
Measurement uncertainty , Uncertainty budget , Gravimetric titration , Micro-Winkler titration , Dissolved oxygen , Primary method
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta