Title of article :
Purity determination and uncertainty evaluation of theophylline by mass balance method, high performance liquid chromatography and differential scanning calorimetry Original Research Article
Author/Authors :
Kang Ma، نويسنده , , Haifeng Wang ، نويسنده , , Min Zhao، نويسنده , , Jinjing Xing، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
Purity determination of pure organic substance is essential for the establishment of traceability to SI units. A mass balance method was employed to determine the purity of theophylline certified reference materials (CRM), compared with high performance liquid chromatography (HPLC) and differential scanning calorimetry (DSC). In the approach of the mass balance, the impurities were identified by ion trap time-of-flight mass spectrometer (IT-TOF-MS) and quantified by HPLC. The purity of theophylline CRM determined by mass balance method was 99.82% with an extended uncertainty of 0.1% (k = 2). The uncertainty evaluation of purity demonstrated that the accuracy of the mass balance method is better than that of HPLC and DSC. It indicated that the mass balance is suitable for the CRM and pharmaceutical standards.
Keywords :
Mass balance , Purity , Uncertainty , Differential scanning calorimetry , Theophylline , Certified reference materials , High performance liquid chromatography
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta