Title of article
Purity determination and uncertainty evaluation of theophylline by mass balance method, high performance liquid chromatography and differential scanning calorimetry Original Research Article
Author/Authors
Kang Ma، نويسنده , , Haifeng Wang ، نويسنده , , Min Zhao، نويسنده , , Jinjing Xing، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
7
From page
227
To page
233
Abstract
Purity determination of pure organic substance is essential for the establishment of traceability to SI units. A mass balance method was employed to determine the purity of theophylline certified reference materials (CRM), compared with high performance liquid chromatography (HPLC) and differential scanning calorimetry (DSC). In the approach of the mass balance, the impurities were identified by ion trap time-of-flight mass spectrometer (IT-TOF-MS) and quantified by HPLC. The purity of theophylline CRM determined by mass balance method was 99.82% with an extended uncertainty of 0.1% (k = 2). The uncertainty evaluation of purity demonstrated that the accuracy of the mass balance method is better than that of HPLC and DSC. It indicated that the mass balance is suitable for the CRM and pharmaceutical standards.
Keywords
Mass balance , Purity , Uncertainty , Differential scanning calorimetry , Theophylline , Certified reference materials , High performance liquid chromatography
Journal title
Analytica Chimica Acta
Serial Year
2009
Journal title
Analytica Chimica Acta
Record number
1037522
Link To Document