• Title of article

    Purity determination and uncertainty evaluation of theophylline by mass balance method, high performance liquid chromatography and differential scanning calorimetry Original Research Article

  • Author/Authors

    Kang Ma، نويسنده , , Haifeng Wang ، نويسنده , , Min Zhao، نويسنده , , Jinjing Xing، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    7
  • From page
    227
  • To page
    233
  • Abstract
    Purity determination of pure organic substance is essential for the establishment of traceability to SI units. A mass balance method was employed to determine the purity of theophylline certified reference materials (CRM), compared with high performance liquid chromatography (HPLC) and differential scanning calorimetry (DSC). In the approach of the mass balance, the impurities were identified by ion trap time-of-flight mass spectrometer (IT-TOF-MS) and quantified by HPLC. The purity of theophylline CRM determined by mass balance method was 99.82% with an extended uncertainty of 0.1% (k = 2). The uncertainty evaluation of purity demonstrated that the accuracy of the mass balance method is better than that of HPLC and DSC. It indicated that the mass balance is suitable for the CRM and pharmaceutical standards.
  • Keywords
    Mass balance , Purity , Uncertainty , Differential scanning calorimetry , Theophylline , Certified reference materials , High performance liquid chromatography
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2009
  • Journal title
    Analytica Chimica Acta
  • Record number