    Indirect pulsed electrochemical detection of amino acids and proteins following high performance liquid chromatography Original Research Article

  • Author/Authors

    Mark P. Olson، نويسنده , , Loryn R. Keating، نويسنده , , William R. LaCourse، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    7
  • From page
    198
  • To page
    204
  • Abstract
    Pulsed electrochemical detection (PED) following liquid chromatographic separation has been applied to the indirect determination of amino acids and proteins. Here, the adsorption of these analytes at noble metal electrodes is exploited to suppress the oxidation of polyols and carbohydrates under alkaline conditions to elicit an indirect response. Of the reagents tested, gluconic acid gave the best overall signal-to-noise values for the indirect detection of amino acids following high performance anion-exchange chromatography (HPAEC). Limits of detection of amino acids were found to be 2–30 pmol using optimized potential–time waveforms at an Au electrode. Indirect PED provided much greater detection sensitivity toward amino acids than direct PED. Analytical sensitivity of indirect PED is a function of both the analyteʹs ability to adsorb to the electrode surface and its molecular size, which was demonstrated by the separation and detection of bovine serum albumin, ovalbumin, and myoglobin following gel-filtration chromatography (GFC).
  • Keywords
    Amino acids , Pulsed electrochemical detection , Indirect detection , High performance liquid chromatography , Proteins