    Downscaling a multicommuted flow injection analysis system for the photometric determination of iodate in table salt Original Research Article

  • Author/Authors

    Sivanildo S. Borges، نويسنده , , Jailson de Souza Peixoto، نويسنده , , M?rio A. Feres، نويسنده , , Boaventura F. Reis، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    5
  • From page
    3
  • To page
    7
  • Abstract
    In this work a downscaled multicommuted flow injection analysis setup for photometric determination is described. The setup consists of a flow system module and a LED based photometer, with a total internal volume of about 170 μL. The system was tested by developing an analytical procedure for the photometric determination of iodate in table salt using N,N-diethyl-henylenediamine (DPD) as the chromogenic reagent. Accuracy was accessed by applying the paired t-test between results obtained using the proposed procedure and a reference method, and no significant difference at the 95% confidence level was observed. Other profitable features, such as a low reagent consumption of 7.3 μg DPD per determination; a linear response ranging from 0.1 up to 3.0 m IO3−, a relative standard deviation of 0.9% (n = 11) for samples containing 0.5 m IO3−, a detection limit of 17 μg L−1 IO3−, a sampling throughput of 117 determination per hour, and a waste generation 600 μL per determination, were also achieved.
  • Keywords
    Iodate , Multicommuted flow analysis , Downsized flow system , Spectrophotometry , Green chemistry
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2010
  • Journal title
    Analytica Chimica Acta
  • Record number