• Title of article

    GC determination of N-nitrosamines by supersonic molecular beam MS equipped with triple quadrupole analyzer, GC/SMB/QQQ/MS Original Research Article

  • Author/Authors

    Voloshenko Anna، نويسنده , , Shelkov Rimma، نويسنده , , Ovadia Lev، نويسنده , , Gun Jenny، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    8
  • From page
    162
  • To page
    169
  • Abstract
    The determination of 14 N-nitrosamines by a supersonic molecular beam electron ionization mass spectrometer equipped with triple quadruple analyzer, GC/SMB/EI/QQQ/MS is presented. The supersonic molecular beam electron ionization ion source allows the elucidation of the molecular ion of 13 out of the 14 examined nitrosamines (except for diphenylnitrosamine which was degraded before the analysis). It was possible to use the molecular ions of all the nitrosamines as the parent ions for multiple reactions monitoring mode, which in turn allows significant increase of specificity and lowering of the method limit of detection of the higher molecular weight nitrosamines. The instrumental LOD for different N-nitrosamines was 1–5 pg injection−1. The proposed method was exemplified by analysis of N-nitrosamines and N-nitrosatables in rubber teats according to the British Standard BS EN 12868:1999.
  • Keywords
    N-Nitrosamines , Soft ionization , Supersonic molecular beam
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2011
  • Journal title
    Analytica Chimica Acta
  • Record number

    1038767