• Title of article

    Development of polymer-membrane based electrodes for suramin Original Research Article

  • Author/Authors

    Andrew Yu، نويسنده , , Brandon Shepherd، نويسنده , , Meghan Wagner، نويسنده , , Jamie Clapper، نويسنده , , Joan M. Esson، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    5
  • From page
    76
  • To page
    80
  • Abstract
    The development of a polymer membrane-based electrode to measure the anionic drug suramin in buffered saline and biological samples is described. A large non-equilibrium, steady state EMF response is observed toward suramin, and judicious choice of the polymer membrane components allows for adjustment of the dynamic range of the electrode. The optimized membrane for use in the toxic suramin range consists of 25 wt% tridodecylmethyl ammonium chloride, 55 wt% bis-2-ethylhexyl sebacate, and 20 wt% Pellethane. Although this electrode can be used to directly quantify suramin in human plasma, determination of suramin that is not affected by the background concentration of small anions is best achieved by simple potentiometric titrations with polycationic protamine monitored with a protamine-sensitive electrode.
  • Keywords
    Suramin , Polymer membrane-based electrode , Potentiometry , Polyanion-sensitive electrode
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2011
  • Journal title
    Analytica Chimica Acta
  • Record number

    1038781