Title of article :
Analysis of energy dissipation in fretting corrosion experiments with materials used as hip prosthesis
Author/Authors :
Kyungmok Kim، نويسنده , , Jean Geringer، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2012
Pages :
7
From page :
497
To page :
503
Abstract :
This paper analyses energy dissipation of fretting corrosion in total hip prosthesis. Fretting corrosion is arisen between metallic prosthesis and bone and/or bone cement, leading to aseptic loosening. In this study, fretting corrosion tests are conducted in Ringerʹs solution. Stainless steel (316L) and poly (methyl methacrylate) are used for total hip prosthesis. Various potentials are applied in fretting corrosion tests and then dissipated energy is determined with number of cycles. Results show that dissipated energy is rapidly accumulated during the initial running-in period and accumulation of dissipated energy change can be expressed with a power-law form. After the initial running-in period, dissipated energy is linearly accumulated with respect to number of cycles. It is identified that a parameter in the power-law relation can describe the influence of applied potentials in fretting corrosion. In addition, the parameter shows relation to wear volume measured in stainless steel.
Keywords :
Fretting wear , Applied potential , Fretting corrosion , Energy dissipation , Total hip prosthesis
Journal title :
Wear
Serial Year :
2012
Journal title :
Wear
Record number :
1092576
Link To Document :
بازگشت