Issue Information
ماهنامه با شماره پیاپی سال 2005
Pages
-902
From page
903
To page
0
Abstract
This article reports on the use of cobalt(II) phthalocyanine (CoPc)-modified carbon paste amperometric detector for monitoring hydrazine compounds following their microchip separation. The marked catalytic electrochemical properties of CoPc-modified electrode display enhanced sensitivity compared with unmodified carbon pastes at a relatively low detection potential (+0.5 V versus Ag/AgCl). Factors influencing the on-chip separation and detection processes have been optimized. Three hydrazines (hydrazine, 1,1 dimethylhydrazine, and phenylhydrazine) have been separated within 130 s at a separation voltage of 1 kV using a 10 mM phosphate run buffer (pH 6.5). The detection limits obtained from using the CoPc-modified carbon paste electrodes for hydrazine and phenylhydrazine are 0.5 and 0.7 (mu)M, respectively, with linearity over the 20200 (mu)M range examined. Such miniaturization and speed advantages of microchip CE are coupled to the highly sensitivity and convenient preparation of CoPc-modified carbon paste electrode. The resulting microsystem should be attractive for field monitoring of toxic hydrazine compounds in environmental applications.
Journal title
Talanta
Serial Year
2005
Journal title
Talanta
Record number
111635
Link To Document