Title of article :
Analysis of the cyclotron facility calibration and aircraft dosimetry results from the liulin-3M instrument Original Research Article
Author/Authors :
Ts.P. Dachev، نويسنده , , E.G. Stassinopoulos، نويسنده , , B.T. Tomov، نويسنده , , Pl.G. Dimitrov، نويسنده , , Yu.N. Matviichuk، نويسنده , , V.A. Shurshakov، نويسنده , , V.M. Petrov، نويسنده ,
Issue Information :
دوهفته نامه با شماره پیاپی سال 2003
Pages :
5
From page :
67
To page :
71
Abstract :
The LIULIN-3M instrument is a further development of the LIULIN dosimeter-radiometer, used on the MIR space station during the 1988–1994 time period. The LIULIN-3M is designed for continuous monitoring of the radiation environment during the BION-12 satellite flight in 1999. A semiconductor detector with 1 mm thickness and 1 cm2 area is contained in the instrument. Pulse high analysis technique is used to determine the energy losses in the detector. The final data from the instrument are the flux and the dose rate for the exposure time and 256 channels of absorbed dose spectra based on the assumption that the particle flux is normal to the detector. The LIULIN-3M instrument was calibrated by proton fluxes with different energies at the Indiana University Cyclotron Facility in June 1997 and had been used for radiation measurements during commercial aircraft flights. The calibration procedure and some flight results are presented in this paper.
Journal title :
Advances in Space Research
Serial Year :
2003
Journal title :
Advances in Space Research
Record number :
1128852
Link To Document :
بازگشت