    Analysis of the cyclotron facility calibration and aircraft dosimetry results from the liulin-3M instrument Original Research Article

  • Author/Authors

    Ts.P. Dachev، نويسنده , , E.G. Stassinopoulos، نويسنده , , B.T. Tomov، نويسنده , , Pl.G. Dimitrov، نويسنده , , Yu.N. Matviichuk، نويسنده , , V.A. Shurshakov، نويسنده , , V.M. Petrov، نويسنده ,

  • Issue Information
    دوهفته نامه با شماره پیاپی سال 2003
  • Pages
    5
  • From page
    67
  • To page
    71
  • Abstract
    The LIULIN-3M instrument is a further development of the LIULIN dosimeter-radiometer, used on the MIR space station during the 1988–1994 time period. The LIULIN-3M is designed for continuous monitoring of the radiation environment during the BION-12 satellite flight in 1999. A semiconductor detector with 1 mm thickness and 1 cm2 area is contained in the instrument. Pulse high analysis technique is used to determine the energy losses in the detector. The final data from the instrument are the flux and the dose rate for the exposure time and 256 channels of absorbed dose spectra based on the assumption that the particle flux is normal to the detector. The LIULIN-3M instrument was calibrated by proton fluxes with different energies at the Indiana University Cyclotron Facility in June 1997 and had been used for radiation measurements during commercial aircraft flights. The calibration procedure and some flight results are presented in this paper.