    Calibration of saturated liquid helium weir flowmeter at Fermilab Original Research Article

  • Author/Authors

    M.W. McGee، نويسنده , , D.M. France، نويسنده , , R.C. Sanders، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    10
  • From page
    759
  • To page
    768
  • Abstract
    The flow of saturated liquid helium at 4.2 K was studied in a unique weir flowmeter mounted inside a duct. The objective was to develop a calibration for helium flow rate in the weir, and thereby, making it a viable flowmeter. The weir flowmeter was tested in a liquid helium system at Fermi National Laboratory used for cooling the superconducting magnets in the experimental areas of the site. Unlike for open channel weir devices, the calibration of this flowmeter was influenced by both the liquid flow through the weir notch and the vapor flow over it. The results of this study and the calibration are presented taking into account effects of both liquid and vapor.
  • Keywords
    Liquid helium , Two-phase helium flow , Weir flowmeter