Title of article
Liquid helium cryostat for SQUID-based MRI receivers Original Research Article
Author/Authors
H.C. Seton، نويسنده , , J.M.S. Hutchison، نويسنده , , D.M. Bussell، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
8
From page
348
To page
355
Abstract
We describe a liquid helium cryostat, developed to cool SQUID-based receivers in low field MRI systems. The cryostat has a 4 L liquid helium capacity, a hold time of over 3 days and accommodates 10 cm diameter receiver coils. New vacuum insulation methods reduce the noise level by at least an order of magnitude compared to existing commercial designs. The minimum detectable field at 425 kHz, with a 5 cm diameter circular coil, was estimated to be 0.018 fT/Hz1/2 from Q-factor measurements and 0.035 fT/Hz1/2 by direct measurement with a SQUID amplifier. Further measurements indicated that most of this field noise probably originates with dielectric losses in the cryostat’s fibreglass shells.
Keywords
Ceramics (A) , Liquid helium (B) , SQUIDs (D) , Cryostats (F)
Journal title
Cryogenics
Serial Year
2005
Journal title
Cryogenics
Record number
1172549
Link To Document