Title of article :
Thermal conductivity of subcooled liquid oxygen Original Research Article
Author/Authors :
D. Celik، نويسنده , , S.W. Van Sciver، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
The thermal conductivity of liquid oxygen below 80 K and pressures up to 1 MPa has been measured using a horizontal, guarded, flat-plate calorimeter. The working equation of the calorimeter is based on the one-dimensional Fourier’s law. The gap between the calorimeter plates was measured in situ from a capacitance measurement. The cooling power to the calorimeter is provided by a two-stage Gifford–McMahan cryocooler. The absolute temperatures are measured using platinum resistance thermometers. The results are compared to existing data and analytical models.
Keywords :
Oxygen (B) , Thermal conductivity (C) , Calorimeters (D)
Journal title :
Cryogenics
Journal title :
Cryogenics