Title of article :
Performance of a helium circulation system for a MEG Original Research Article
Author/Authors :
Tsunehiro Takeda، نويسنده , , Masayoshi Okamoto and Nobuyuki Shima ، نويسنده , , Kazuhiro Atsuda، نويسنده , , Keishi Katagiri، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
We report the performance of a helium circulation system (HCS) for a magnetoencephalography (MEG) that re-liquefies all the evaporating helium gas using two 1.5 W GM cryocoolers operating at 4.2 K. The MEG with the HCS was used to measure human brain responses for over one and a half years without any noise problems. The noise level is below 10 fT/Hz1/2 for 2–40 Hz, below 30 fT/Hz1/2 at 1 Hz, and 200 fT/Hz1/2 at 50 Hz, which is the power supply frequency. As the amount of liquid helium used decreases less than one percent, the maintenance cost of the MEG becomes less than one-tenth of the previous cost.
Keywords :
SQUIDs (D) , Gifford–McMahon (E) , Cryostats (F) , Reuse (B) , Liquid helium (B)
Journal title :
Cryogenics
Journal title :
Cryogenics