• Title of article

    Amperometric enzyme electrode for free cholesterol determination prepared with cholesterol oxidase immobilized in poly(vinylferrocenium) film

  • Author/Authors

    B. Cem ?zer، نويسنده , , Haluk ozyoruk، نويسنده , , Serdar S. Celebi، نويسنده , , Atilla Y?ld?z، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    4
  • From page
    262
  • To page
    265
  • Abstract
    The steady state amperometric measurements of free cholesterol were performed by an enzyme electrode which was developed through electrostatic immobilization of the cholesterol oxidase in poly(vinylferrocenium) perchlorate, PVF+ClO4−, film that was coated on a Pt electrode. Amperometric measurements were performed at an applied potential of 0.7 V versus SCE. The effects of enzyme concentration, polymer film thickness, working temperature, dissolved oxygen gas and pH of solution on the response electrode were elucidated. The activation energy and the apparent Michaelis–Menten constant were calculated for enzyme electrode system. The linear working range, the sensitivity and the stability of the biosensor were also determined. The mechanism for the operation of the enzyme electrode was proposed.
  • Keywords
    Poly(vinylferrocenium) , Cholesterol oxidase , Amperometric enzyme electrode , Free cholesterol biosensor
  • Journal title
    Enzyme and Microbial Technology
  • Serial Year
    2007
  • Journal title
    Enzyme and Microbial Technology
  • Record number

    1174824