Title of article
Determination of Michaelis–Menten parameters obtained from isothermal flow calorimetric data
Author/Authors
M.A.A. O’Neill، نويسنده , , A.E Beezer، نويسنده , , J.C. Mitchell، نويسنده , , J Orchard، نويسنده , , J.A. Connor، نويسنده ,
Issue Information
دوهفته نامه با شماره پیاپی سال 2004
Pages
6
From page
187
To page
192
Abstract
Recent papers have reported [Thermochim. Acta 399 (2003) 63; Thermochim. Acta, in press] the results of a preliminary inter/intra laboratory study into the suitability of the base-catalysed hydrolysis of methyl paraben as a test and reference reaction for isothermal flow-through calorimeters. It was shown that this reaction can be used to investigate the flow characteristics of the instrument being used. It has also allowed, for the first time, the calculation of accurate values for the rate constant and for the enthalpy change, ΔH (hereafter H (enthalpy) for simplicity) of reaction directly from the calorimetric data, free from assumption. These findings have been extended to permit the direct determination of Michaelis–Menten based kinetic parameters from calorimetric data again free from assumption (except that the system conforms to Michaelis–Menten kinetic theory). This paper describes the method used for such an analysis and reports the results of a preliminary study on the urea/urease enzymatic system.
Keywords
Michaelis–Menten kinetics , Enzymes , Flow calorimetry
Journal title
Thermochimica Acta
Serial Year
2004
Journal title
Thermochimica Acta
Record number
1196435