Title of article
Subtracting the water effect from DSC curves by using simultaneous TGA data
Author/Authors
Ramon Artiaga، نويسنده , , Salvador Naya، نويسنده , , A. Garc?a، نويسنده , , F. Barbadillo، نويسنده , , Laura Garc?a، نويسنده ,
Issue Information
دوهفته نامه با شماره پیاپی سال 2005
Pages
3
From page
137
To page
139
Abstract
In this work, a method to calculate the baseline when weight loss processes take place is presented. Such method combines the information obtained by TGA with the simultaneous heat flow signal obtained in a Simultaneous Thermal Analyzer. Once the water evaporation step is identified, thermal parameters such as specific heat and enthalpy of vaporization can be combined with the instantaneous weight loss to subtract the heat flow corresponding to this component from the DSC baseline. In this paper the model was applied to water evaporation in forest litter samples.
Keywords
Simultaneous DSC TGA , Subtraction
Journal title
Thermochimica Acta
Serial Year
2005
Journal title
Thermochimica Acta
Record number
1196746
Link To Document