Title of article :
Calibration and desmearing of a differential thermal analysis measurement signal – upon heating and cooling – in the high-temperature region
Author/Authors :
Wolfgang Baumann، نويسنده , , Andreas Leineweber، نويسنده , , Eric Jan Mittemeijer، نويسنده ,
Issue Information :
دوهفته نامه با شماره پیاپی سال 2008
Pages :
5
From page :
50
To page :
54
Abstract :
A previously reported calibration and desmearing method for a differential thermal analysis (DTA) apparatus applicable to heating and cooling was successfully modified to be used in the high-temperature region (up to 1570 K). The method requires knowledge of parameters, which have a physical meaning, of the heat-flux model appropriate for the DTA apparatus. Values for these physical parameters were determined by calibration. As calibration materials Mo (exhibiting a smooth variation of specific heat with temperature) and Co (non-monotonic change of specific heat upon the ferro- to paramagnetic transition and vice versa) were employed. Experimental measures to improve the reproducibility were presented.
Keywords :
Desmearing , DTA , Cobalt , Curie point , High-temperature region , calibration
Journal title :
Thermochimica Acta
Serial Year :
2008
Journal title :
Thermochimica Acta
Record number :
1198160
Link To Document :
بازگشت