Title of article :
Position determination of interaction points inside a planar-type segmented germanium detector by pulse shape analysis using a three-dimensional Greenʹs function Original Research Article
Author/Authors :
J.Y. Moon، نويسنده , , C.S. Lee، نويسنده ,
Issue Information :
هفته نامه با شماره پیاپی سال 2004
Pages :
4
From page :
643
To page :
646
Abstract :
Pulse shape analysis was performed to determine three-dimensional positions of interaction points inside a planar-type segmented germanium gamma-ray detector. The crystal has surface dimensions of 5 cm × 5 cm with a thickness of 2 cm and is segmented into a total of 25 segments. Pulse shapes of a hole and an electron, respectively induced on each segment (cathode) and an anode, were analytically obtained in 1-ns intervals by using the three-dimensional Greenʹs function method. For the depth position determination, we used the two-dimensional rise-time correlation between the anode and cathode signals.
Journal title :
Nuclear physics A
Serial Year :
2004
Journal title :
Nuclear physics A
Record number :
1201930
Link To Document :
بازگشت